Wash-Down Fume Cupboard
Wash-Down Fume Cupboard
Wash-Down Fume Cupboard

Verified 3 months ago

Tion Global Ltd

Verified 3 months ago

Tion Global Ltd

Verified 3 months ago

Tion Global Ltd
Summary
A fume cupboard designed to prevent explosive perchlorate salt formation.
-
-
Application

Suitable for a range of research, pharmaceutical and military applications.

-
-
Description

TION’s wash-down fume cupboard range is a ducted, high-capacity product designed to eliminate perchloric salt formation. The unit is equipped with a touch-screen control system and airflow alarms to maintain your staff’s safety no matter the circumstance.


Wash-down fume cabinets are fixed fume extraction systems fitted with angular walls and bespoke water spray nozzles. These nozzles wash chemical residues from behind the rear baffles and duct connections, eliminating the chance of explosive perchloric salt formation inside your laboratory.


Features and benefits:

  • Corrosion-proof build.
  • BS EN 14175 and ASHRAE 110-1995 certified.
  • Customisable dimensions.
  • Full installation and commissioning.
  • Integrated water sprinklers.
  • Work surface options include polypropylene, stainless steel, ceramic and epoxy.
  • Red-light alarm system.
  • 100% polypropylene carcass.
  • 7-inch (standard model) or 9-inch (pro model) touch-screen.
  • Variable air volume available on the 9-inch pro panel.
-
-
Digital objects
-
-
Color
White
-
-
Finish
Other
-
-
Material
Polypropylene
-
-
Shape
Rectangle
-
-
Length
1200 mm, 1500 mm, 1800 mm, 2000 mm, 2500 mm, Bespoke
-
-
Width
980 mm, Bespoke
-
-
Height
2370 mm, Bespoke
-
-
Warranty description
One year
-
-
SKU Code
PPFC1200-WD, PPFC1500-WD, PPFC1800-WD, PPFC2000-WD, PPFC2500-WD
-
-

Specification

Standard
To BS EN 14175.
-
-
Arrangement
Free standing unit.
-
-
Extract air
-
-
-
Face velocity
0.5 ±0.1 m/s (100 ±20 FPM).
-
-
Operation
Facility to vary extract volume flow rate.
-
-
Materials
Polypropylene.
-
-
Sash
-
-
-
Sash arrangement
Vertical., Horizontal., Vertical/ Horizontal.
-
-
Operation
Powered, automatic.
-
-
Maximum operational sash opening
500 mm.
-
-
Dimensions
-
-
-
Overall depth
Bespoke.
-
-
Overall width
Bespoke.
-
-
Lighting
-
-
-
Lighting type
LED lighting.
-
-
Illuminance at work surface (minimum)
800 lx.
-
-
Services
-
-
-
Water
-
-
-
Sink unit
Not required., Required.
-
-
Water supply
Not required., Cold water supply., Hot water supply., Reverse osmosis water supply., Demineralized water supply.
-
-
Electricity
-
-
-
Power requirements
110 V, 50/60 Hz, single-phase., 220 V, 50/60 Hz, single-phase.
-
-
Location
Sockets on front face of fume cupboard.
-
-
Gas outlets
Not required., Nitrogen., Argon., Hydrogen., Vacuum., Compressed Air., Other.
-
-
Airflow indicator with audible and visual alarms
Constant airflow monitoring and display with visual alarm on screen and integrated internal red light alarm system.
-
-
Controls
-
-
-
Controller type
Touch screen control panel, 9-inch.
-
-
Control modes
Eco mode.
-
-
Accessories
Not required., Glove apertures., Multisash., Partitions., Shelves., Sumps., Vacuum tap.
-
-

Additional properties

Worksurface
HPL., Ceramic., Epoxy., Polypropylene., Stainless steel.
-
-
Safety devices
Not required., Co2/ powder fire suppression., External interlock emergency stop.
-
-

Sustainability data

Contains Red List materials
No
-
-
Country of material origin
Germany
-
-
Material
Polypropylene
-
-
Expected Life
Ten year
-
-