Verified by manufacturer 2 months ago
PPFC120– Standard Model, Constant Air Volume
ISO 14001 Environmental Management Systems
Certificate type
ISO 9001 Quality Management
Certificate type
Alcumus SafeContractor
Certificate type
A metal-free, corrosion-proof fume extraction solution designed for chemical-intensive applications.
Suitable for a range of research, educational, pharmaceutical, oil and gas, industrial and military applications.
The Ducted Fume Cupboard is a medical-grade, welded polypropylene fume cupboard, purpose-built for general and acid-based processes. Ducted fume cabinets capture hazardous aerosols and vapours generated during the scientific process. The build materials ensure that there is no risk of corrosion or degradation over time from a wide range of chemicals.
Features and benefits:
Color | White |
Finish | Other |
Material | Polypropylene |
Shape | Rectangle |
Length | 1200 mm 1500 mm 1800 mm 2000 mm 2500 mm |
Width | 830 mm Bespoke |
Height | 2320 mm Bespoke |
Warranty description | One year |
Uniclass |
Standard | To EN 14175. |
Arrangement | Free standing unit. |
Special requirements | Acid-based processes. |
Extract air | |
Face velocity | 0.5 ±0.1 m/s (100 ±20 FPM). |
Operation | Continuous. |
Materials | Polypropylene. |
Sash | |
Sash arrangement | Horizontal. |
Operation | Manual. |
Maximum operational sash opening | 500 mm.Guidance for specification option: Maximum sash height 720 mm (unless otherwise specified). |
Dimensions | |
Overall depth | 830 mm.Bespoke.Guidance for specification option: Insert requirements. |
Overall width | 1200 mm. |
Lighting | |
Lighting type | LED lighting. |
Illuminance at work surface (minimum) | 800 lx. |
Services | |
Water | |
Sink unit | Not required.Required. |
Water supply | Not required.Cold water supply.Hot water supply.Reverse osmosis water supply.Demineralized water supply. |
Drainage | Manufacturer guidanceInsert requirements or omit. |
Electricity | |
Power requirements | 110 V, 50/60 Hz, single-phase.220 V, 50/60 Hz, single-phase.Manufacturer guidanceA three-phase power supply can be ordered. Consult manufacturer for details. |
Electrical socket outlets | Four.Manufacturer guidanceSingle phase, 3A, 240 V, 50 Hz. |
Location | Sockets on front face of fume cupboard. |
Gas outlets | Not required.Nitrogen. |
Airflow indicator with audible and visual alarms | Constant airflow monitoring and display with visual alarm on screen and integrated internal red light alarm system. |
Controls | |
Controller type | Touch screen control panel, 7-inch.Guidance for specification option: Full-colour touch screen with air flow velocity display, low airflow alarms display, fan control on/off and calibration settings. |
Control modes | Eco mode, Night setback mode (pro only). |
Accessories | Not required.Centrifugal fans.Glove apertures.Multisash.Partitions.Shelves.Sumps.Vacuum tap. |
Worksurface | Polypropylene. Stainless steel. Ceramic. Epoxy. Trespa. |
Safety devices | Not required. Co2/powder fire supression. External interlock emergency stop. |
Contains Red List materials | No |
Country of material origin | Germany |
Material | Polypropylene |
Expected Life | Ten year |
NBS BIM Object Standard
This BIM object has been authored to meet the requirements of the NBS BIM Object Standard.
Find out more about the NBS BIM Object Standard.
Platform | Compatible | Version |
---|---|---|
- | 1 | |
- | 1 |