NBS Source
I'm a manufacturer

Application

Suitable for use in operating theatres.

Description

Brandon Medical offer a range of Emergency Power Units (iBEPU) designed to provide a single operating light with a maintained 24 V d.c. supply for a period of no less than 3 hours, in accordance with the regulatory and recommendatory documentation.


The iBEPU is supplied via 230 V a.c., which is utilised to provide 24 V d.c. to power the lamp under normal (healthy) operating conditions and charge the batteries. Should the primary supply suffer a mains failure or a catastrophic component failure, then the iBEPU will switch over to its own battery supply.


The monitored iBEPU system provides the additional benefit of indicating fault conditions of the charger or batteries, that could otherwise remain undetected until there is a power outage.


Features and benefits:

  • An integrated power supply unit (PSU).
  • A separate battery charger.
  • 10-year design life batteries.
  • High speed (static switch) changeover unit.
  • Intelligent theatre control panel (iTCP) connection (via Modbus communication).
  • Local or external indicators.
  • Includes Near Field Communication (NFC).
  • Collects data for secure reports via Android app.
  • Available in wall or panel mount versions for flexibility. The panel mount version must be housed in a suitable enclosure.

General information

Color

White

Material

Steel

Guidance for specification option:

Enclosure.

Shape

Rectangle

Length

400 mm

Width

200 mm

Height

400 mm

Uniclass

Pr_60_70_06_28 Emergency call battery backup supplyPrimary

Specification data - Emergency call battery backup supply

Standards

EN60601-1-2:2015, EN60601-1:2006+A12:2014, HTM06-01.

Battery type

2 x 12V 18Ah in series.

Duration of operation

Three hours.

Battery enclosure

Steel, painted RAL 9010.

Input voltage

100–240 V a.c.

Voltage

20–28.5 V d.c.

Current

Output Current: 4 A (maximum). Discharge current: 30 Amps (maximum).

Ingress protection (IP) rating

IP20.

BIM objects

BOS

NBS BIM Object Standard

This BIM object has been authored to meet the requirements of the NBS BIM Object Standard.

Find out more about the NBS BIM Object Standard.

Platform Compatible Version
Revit
-

1

IFC
-

1

Literature

Declaration of Conformity EU MDR_iBEPU

Declaration of Conformity EU MDR_iBEPU

TB029 iBEPU_iTCP RS485 Connection guidance

TB029 iBEPU_iTCP RS485 Connection guidance

IG-iBEPU [B1] iBEPU INSTALLATION GUIDE

IG-iBEPU [B1] iBEPU INSTALLATION GUIDE

OM-iBEPU [B5]

OM-iBEPU [B5]