NBS Source
I'm a manufacturer
Compare variants

Medipower™ Triple Cabinet IPS - Medical IT Power Supply

Verified by manufacturer 5 months ago

Transformer Size (Rated Power): 4 kVA

Brandon Medical Co Ltd Variant:

Isolated power supply for group 2 medical locations.

Jump to:
BOS
+44 (0)113 277 7393 Website

Application

Suitable for HD60364-7-710 Group 2 locations as well as HTM06-01 Category 5 areas.


Medical IPS systems should serve the following areas:

  • Operating theatre suites and recovery areas.
  • Emergency departments – resuscitation.
  • Critical care areas – CCU, ITU, SCBU, NICU, HDU etc.
  • Imaging rooms – MRI, CT, PET, X-Ray, Cath Lab etc.

Description

Medical IT Power Supply system (IPS) provides safe delivery of electrical power. The Medipower Medical IT Power Supply system (IPS) provides continuity of supply in the instance of the first insulation fault and also offers additional protection from electrical leakage currents, for group 2 locations or HTM06-01 for clinical risk category 5 areas.


Features and benefits:

  • High quality Medipower Medical IT Power Supply system (IPS) ensures safe energy distribution.
  • Distinct monitoring and distribution panel incorporates both wall-mounted and floor-standing options within the same panel design.
  • Automatic electrical supply, the standards state that Group 2 Medical Locations shall have an automatic electrical supply available within 0.5 seconds of power failure. In the UK, it is commonplace to support the IPS system with an upstream Uninterruptible Power Supply (UPS) in order to meet the 0.5-second requirement. In other parts of the world, this may be by switching to an alternative source of supply.
  • A CANbus (Controller Area Network bus) allows devices to communicate with each other without a host computer. IPS alarm signals communicate seamlessly via CANbus, offering many benefits.
  • CANbus is capable of identifying and reporting load insulation faults in 5–30 seconds.
  • Each CANbus network can be up to 2500 m in length for linking multiple networks or future expansion.
  • CANbus can offer over 50% reduction in field cabling costs.
  • CANbus continually monitors two-way communications between all CANbus connected devices, identifies any faults and provides parallel redundancy.
  • Modular construction and powerful messaging.
  • Ability to hot-swap and maintain individual components. Programmable content with up to 1000 alarm messages.
  • Smaller footprint space requirements.
  • IPS units can be supplied in as single (SIPS), double (DIPS) or triple (TIPS) configurations with a variety of different transformer sizes and in IPS, IPS+IFS or IPS+IFS+BATS variants.
  • Flexible number of outgoing ways - IPS system MCBs are available in functional blocks of six, up to a maximum of 96.
  • Multiple master alarm functionality.
  • Advanced event logging.

General information

Color

Grey

Guidance for specification option:

RAL 7035.

Finish

Powder-coated

Material

Steel

Shape

Rectangle

Length

350 mm

Width

800 mm

Height

1700 mm

Uniclass

Pr_60_70_64_51 Medical isolated power supply units (IPS)Primary

Specification data - Medical isolated power supply units (IPS)

Standards

To HD60364-7-710, group 2. To HTM06-01, category 5 areas.

Accessories

Not required.

Pass switch.

Surge protection.

MCB tripped monitoring.

Network gateway.

MODbus interface.

Primary side outgoing RCBO’s (Type – 2C,25).

Configuration

Triple cabinet.

Components

Incoming isolator.
Control and monitoring device - ILT.

Dimensions

1700 x 800 x 350 mm.

Guidance for specification option:

Distribution panel enclosure height / TRX enclosure height: 1100 / 600 mm.

Manufacturer guidance

Height x width x depth.

Enclosure

Mild steel, RAL 7035.

Mounting

Floor.

Wall.

Manufacturer guidance

Distribution panel mount type:

(TRX enclosure always floor mounted).

Fans

3.

Air flow

42 m³/hr per IPS channel.

Ingress protection (IP) rating

IP41.

Protective earth bars

100 A.

Circuit breakers

24.

Manufacturer guidance

MCBs (max. quantity).

Transformers

4 kVA.

Rated frequency

50–60 Hz.

Input voltage

230 V AC.

Current rating

17.4 A.

Manufacturer guidance

Rated Output Current.

Power dissipation

Heat power dissipation no-load: <36 W.
Heat power dissipation 100%-load: <138 W.

Efficiency

96%.

Weight

52 kg.

Guidance for specification option:

Approximate.

Supply

Typical MCB Supply (A): C32 or D25.
Typical MCCB Supply (A): AF63/ AT30.

Sustainability data

Country of product manufacture

United Kingdom

BIM objects

BOS

NBS BIM Object Standard

This BIM object has been authored to meet the requirements of the NBS BIM Object Standard.

Find out more about the NBS BIM Object Standard.

Platform Compatible Version
Revit
-

1

IFC
-

1